摘要: |
[摘要] 目的 分析超声弹性成像对校正乳腺超声影像报告数据系统(Breast Imaging and Reporting System Ultrasound,BI-RADS-US)3类及4类乳腺肿块的诊断价值。方法 对常规BI-RADS-US分类法诊断为3类及4类的乳腺肿块用超声弹性成像进行校正,以病理结果为金标准,对比分析校正前后的BI-RADS-US分类法对乳腺肿块良恶性的鉴别诊断效能。结果 常规BI-RADS-US分类法诊断乳腺良恶性肿块的灵敏度、特异度、准确度分别为75.6%、88.1%、83.8%,受试者工作特征(ROC)曲线下面积为0.836。经超声弹性成像校正后BI-RADS-US分类法的诊断灵敏度、特异度、准确度分别为82.9%、93.1%、90.8%,ROC曲线下面积为0.918。校正前后ROC曲线下面积比较差异有统计学意义(P<0.05)。结论 弹性成像有助于提高超声对BI-RADS-US 3类及4类乳腺肿块良恶性的鉴别诊断效能。 |
关键词: 超声弹性成像 校正 乳腺肿块 乳腺超声影像报告数据系统 |
DOI:10.3969/j.issn.1674-3806.2020.06.14 |
分类号:R 445.1 |
基金项目: |
|
Value of BI-RADS-US rectified by ultrasound elastography in diagnosis of type 3 and type 4 breast lumps |
XI Jian-wu, ZHOU Yi-wei, YE Jian-qiang
|
Department of Medical Ultrasonics, Wuxi Huishan District People′s Hospital, Jiangsu 214187, China
|
Abstract: |
[Abstract] Objective To analyze the value of Breast Imaging and Reporting System Ultrasound(BI-RADS-US) rectified by ultrasound elastography in diagnosis of type 3 and type 4 breast lumps. Methods Breast lumps of type 3 and type 4 diagnosed by conventional BI-RADS-US classification were diagnosed by ultrasound elastography as rectification. With the pathological results as the gold standard, the efficacies of BI-RADS-US classification were compared and analyzed in differential diagnosis of benign and malignant breast tumors before and after rectification. Results The sensitivity, specificity and accuracy of the conventional BI-RADS-US classification were 75.6%, 88.1% and 83.8% respectively in diagnosis of benign and malignant breast tumors, and the area under the receiver operating characteristic(ROC) curve was 0.836. The sensitivity,specificity and accuracy of the BI-RADS-US classification with ultrasound elastography rectification were 82.9%, 93.1% and 90.8% respectively, and the area under ROC curve was 0.918. There were significant differences in the areas under ROC curve before and after rectification(P<0.05). Conclusion Elastography is conducive to improving the diagnostic efficacies of ultrasound in the differential diagnosis of benign and malignant breast tumors of BI-RADS-US type 3 and type 4. |
Key words: Ultrasound elastography Rectification Breast lumps Breast Imaging and Reporting System Ultrasound(BI-RADS-US) |